Skip to main content
The TAG@IARCTrace Analysis Group
Facilities / Positive Pressure Laboratory

Positive Pressure Laboratory

Clean-space infrastructure for low-level environmental measurements, method development, and high-resolution mass spectrometry.

Facility commissioning

Building verified capability in stages.

The PPL is home to a Thermo Scientific Element XR high-resolution inductively coupled plasma mass spectrometer acquired through NSF Major Research Instrumentation award 2511702. The system will expand UAF capacity for precise elemental concentration and isotope-ratio measurements at low concentrations.

Element XRhigh-resolution sector-field ICP-MS
2026–2029award project period
Staged accessmethods released after verification

Draft 0.3 · August 19, 2026

Planned analytical offerings

The PPL intends to develop four named, matrix-specific methods rather than advertise a universal trace-elements analysis. The draft menu below identifies the samples, focused analyte suites, preparation routes, and analytical depth the facility is being built to support.

Concrete targets, not current services. Every offering below is presently proposed. A method will move to pilot and then verified status only after local performance, quality-control rules, sample requirements, capacity, price, and turnaround have been established. The menu will be revised as commissioning produces evidence.
ProposedDefined commissioning target; no public sample submission.
PilotLimited collaborative batches under active evaluation.
VerifiedDocumented method with published submission and reporting requirements.

Analytical depth

Four tiers of work

The method code defines the matrix and analytes. The tier defines the preparation, quality control, review, and reporting supplied for a project. Available tiers will be published separately for each verified method.

Tier 1

Screening survey

Purpose: exploratory elemental surveys, sample triage, and method scouting.

Typical input: approved, instrument-ready dilute-acid solutions or digests.

Planned output: concentration estimates, detected/not-detected flags, and an abbreviated batch-control summary. Screening results would not carry the full claims of a verified quantitative method.

Tier 2

Standard quantitative

Purpose: routine research samples that fit a verified matrix-specific method and reporting range.

Planned controls: multi-point calibration, second-source verification, method blank, appropriate reference material or matrix spike, duplicate, continuing calibration checks, and documented acceptance rules.

Planned output: reviewed concentrations, units, qualifiers, method reference, and batch quality-control summary.

Tier 3

Enhanced quantitative

Purpose: demanding matrices, lower target concentrations, or projects requiring stronger project-specific evidence.

Possible additions: matrix separation or preconcentration, replicate preparation, additional reference materials and spikes, isotope or resolution confirmation, narrower contamination controls, and a project-defined uncertainty assessment.

Access: scoped with facility staff before sample collection.

Tier R

Collaborative research

Purpose: unsupported matrices or analytes, new preparation chemistry, isotope-ratio work, speciation, coupled techniques, and other method development.

Structure: a jointly planned research activity with explicit development time, controls, decision points, data interpretation, and funding. A successful project may later become a standard offering.

Initial portfolio

Four methods being developed

Draft Positive Pressure Laboratory offerings
Code and statusOfferingFocused development panelCandidate matrices and preparationLikely tiers
TAG-TE01
Proposed · Priority 1
Trace elements in waters and dilute extractsFocused subset of the EPA 200.8 analyte suite; mercury is excluded from the initial XR panelFreshwater, river water, snow or ice melt, permafrost leachates, and dilute acid extracts that meet verified turbidity, salinity, and dissolved-solids limits; filtered or total-recoverable claims will be method-specific1, 2, 3
TAG-TE3A
Proposed · Priority 2
Acid-extractable elements in sediments, soils, and particlesFocused subset of EPA 6020B demonstrated analytes; major matrix elements and mercury are excluded from the initial panelMarine or freshwater sediment, suspended particles, soils, and permafrost solids; microwave-assisted nitric acid or nitric/hydrochloric-acid extraction informed by EPA 3051A2, 3
TAG-RE01
Proposed · Priority 3
Rare earth elements plus yttriumY and the lanthanide series, with Th and U evaluated as optional additionsAcid extracts and compatible sediment, soil, permafrost, or particle digests; aqueous matrices only after matrix-specific verification2, 3, R
TAG-SP01
Proposed · Priority 4
Phosphorus and sulfurP and SSelected biological, sediment, soil, permafrost, or particle digests and low-matrix aqueous samples; not a replacement for routine dissolved nutrient analysis2, 3, R

Panels are development targets. Final reportable elements, isotopes, resolution, preparation, working range, detection capability, sample volume, preservation, and exclusions will be method-specific. The detailed Element XR page records the current candidate analytes without crowding this overview.

EPA-informed is not EPA-certified. TAG-TE01 uses the analyte scope of EPA Method 200.8 as a practical starting point. TAG-TE3A draws on the extraction concept in EPA SW-846 Method 3051A and the demonstrated ICP-MS analytes in Method 6020B (PDF). The TAG methods will use the Element XR and locally documented procedures; regulatory equivalency, accreditation, and compliance suitability must not be assumed.

Read the detailed Element XR method-development roadmap

What changes as methods mature

The stable parts

Method codes identify a scientific claim and matrix family. They allow proposals, documentation, data systems, and future publications to refer to the same offering even as its version advances.

Every released method will state exactly which preparation, analytes, matrices, and tiers it supports.

The versioned parts

Panels, reporting ranges, detection capability, sample requirements, batch size, price, turnaround, quality-control limits, and availability can change as evidence and capacity improve. Material changes will receive a new method version and date.

Projects outside a released scope remain Tier R rather than being forced into an unsuitable routine method.

Outside the initial routine menu

Biological-tissue panels, total silicate digestion, dissolved trace metals in seawater or saline brines, total mercury, seawater rare earth elements, isotope ratios, chromatography and speciation, laser ablation, and other coupled applications are not part of the four-method initial portfolio. They may be considered as Tier R collaborations when the necessary preparation, standards, interfaces, expertise, and project support are available.

Planning a future project

Discuss projects with the team before writing a proposal, collecting samples, or promising analyses. Early conversations should cover the scientific question, requested method and tier, target analytes, expected concentrations, matrix, salinity or dissolved solids, available sample volume, preparation and preservation, contamination control, reference materials, number of samples and controls, data requirements, schedule, and funding.